Prolyte USA
I-Typical Within Run (n=30) Blood, Serum, Plasma 2-Lithium is not typically measured in urine samples
3-(60-120) requires additional dilution
Description
ProLyte®
The World’s Most Popular Electrolyte Analyzer
Analysis Time
45 sec (240/hour)
Data Storage
20,000+ Patient Results
10,000+ QC Results per Level
User Input
4″ Touch Screen
External Keyboard
Remote Access
Connectivity
LAN, Wifi, 4 USB ports, 88232, LIS
Calibration
2 Point every 4 hours
1 Point every sample
On Demand
Number ql Parameters
Na’ K’ CI Li‘
4 Total /3 (Simultaneously)
Support
Online Service Interface
Sample Size ‘.
Whole Blood, Serum, Plasma, QCs (100)
Urine (350)
| Parameter | Range | Reproducibility¹ | Resolution |
| Na⁺ | 45–205 mmol/L | CV ≤ 1% (120–160 mmol/L) | 0.01 mmol/L |
| K⁺ | 1.5–15 mmol/L | CV ≤ 1.5% (2.5–6 mmol/L) | 0.01 mmol/L |
| Cl⁻ | 45–205 mmol/L | CV ≤ 1% (85–130 mmol/L) | 0.01 mmol/L |
| Li⁺ | 0.15–5.00 mmol/L | SD ≤ 0.03 mmol/L (0.4–1.3 mmol/L) | 0.001 mmol/L |
| Urine² Na⁺ | 30–1020 mmol/L | CV ≤ 5% (100–250 mmol/L) | 0.01 mmol/L |
| K⁺ | 20–505 mmol/L³ | CV ≤ 5% (10–60 mmol/L) | 0.01 mmol/L |
| Cl⁻ | 25–505 mmol/L | CV ≤ 5% (100–250 mmol/L) | 0.01 mmol/L |
I-Typical Within Run (n=30) Blood, Serum, Plasma 2-Lithium is not typically measured in urine samples
3-(60-120) requires additional dilution
| Power | Size 8: Weight | Ambient Conditions |
| 100—240V ~ 50/60 Hz (self adjusting) | 9.5″ {24cm} W x16.5″ (42cm) | Room temperature: 15-32°C/60—90°F |
| 1.6Amax, 50Watts max | 14 lbs or 6kgs | Humidity <85% |



Reviews
There are no reviews yet.